石墨炉原子吸收分光光度法测定固体废物中的钡

Determination of Barium in Solid Wastes by Graphite Furnace Atomic Absorption Spectrophotometry

  • 摘要: 研究了石墨炉原子吸收分光光度法测定固体废物中钡的方法。确定了石墨管类型、最佳升温程序和酸消解体系等的测定条件,并通过实际样品检测验证了方法的可行性。结果表明,采用热解涂层石墨管,当溶液钡浓度在0~50.0 μg/L,最佳灰化温度为1 000 ℃,最佳灰化时间为8.0 s,最佳原子化温度为2 600 ℃,最佳原子化时间为2.8 s时,钡浓度与吸光度呈良好的线性关系,相关系数为0.999 1,方法检出限为2.7 mg/kg,相对标准偏差为2.7%~5.4%,实际样品的加标回收率为89.5%~94.5%。

     

    Abstract: A method for determination of barium in solid wastes by graphite furnace atomic absorption spectrophotometry was studied. The optimal measurement conditions of the graphite tube type, heating program and acid digestion system were determined, and the feasibility of this method validated by test of actual samples. The results showed that with pyrolytic coated graphite tube and at the ashing temperature of 1 000 ℃, the ashing time of 8.0 s, the atomization temperature of 2 600 ℃ and the atomization time of 2.8 s, the concentration of barium showed a linear relationship with the absorbance, with the correlation coefficient of 0.999 1. The detection limit of barium was 2.7 mg/kg, the relative standard deviation was between 2.7% and 5.4%, and the recovery rate of the actual samples was between 89.5% and 94.5%.

     

/

返回文章
返回